1 - 10
(5)
After completing the connections of the tubes, attach the fixing
plate of the reducing agent mixing tube assembly with the
screws. Then, circulate water through the tubes to check that
there is no leakage.
CAUTION
Be careful about water leakage. An incomplete
connection of the pump tubes may cause liquid
leakage from a connection point. Be sure to
securely tighten the screws of the nipples.
Fig. 1-10 Attachment of Fixing Plate of Tube Assembly
(6) Attach the cover and fix it with the screws.
1.5.3.2 Setting of Pump Tubing
(1) Raise the support clamp lever and put the tube support down.
(2) Connect the four suction tubes provided as accessories, sample
suction tube (part No. 7J1-7300), hydrochloric acid suction tube
(part No. 7J1-7305), sodium borohydride suction tube (part No.
7J1-7310) and reducing agent suction tube (part No. 7J1-7315),
to the four tubes extending from the HFS-4 main unit,
respectively. Each of the tubes extending from the main unit is
marked with a color as shown below for identification.
HCI: red
NaBH
4
: green
Reducing agent: blue
Sample: none