5.3 Troubleshooting
If a trouble occurs on the instrument, it should be eliminated with
reference to Table 5-1. For a trouble outside the range of this table,
contact the local service representative.
Table 5-1 Troubleshooting
Symptoms
Causes (check methods)
Remedies
1. Indicator lamp does not
light after power on.
Pump does not operate.
The power cord at rear is not connected.
(Check the power connecter.)
Connect the power cord
Fuse is not set.
(Check the fuse.)
Be sure to disconnect power
cord, and then set fuse.
Fuse is blown.
(Check the fuse.)
Be sure to disconnect power
cord and then set fuse.
Argon gas is not supplied.
(Check argon gas connection.)
Supply argon gas
Argon gas pressure is inadequate.
(Check gas pressure.)
Adjust gas pressure to
500 kPa.
The argon gas cylinder is empty.
(Check the residual pressure in the
cylinder.)
Replace it with a full
cylinder.
2. Reagent cannot be
aspirated.
(while pump is
operating)
Tube support is not set.
(Check tube support.)
Set tube support firmly.
Pump tube is separated from rotor
groove.
(Check tube length.)
Adjust the length of pump
tube and set the tube again.
Pump tube is torn.
(Check tube.)
Replace the tube.
Tip of reagent tube is not dipped into
each solution.
(Check solution volume and tube tip.)
Dip the tip of tube into
reagent.
Increase solution volume.
Pump tube is hardened due to long
storage at low temperature.
(Aspirate reagent by pressing the tube
support.)
Wait until temperature rises.
Replace the pump tube.
The connection of each tube is loose.
Connect firmly.
Clogged air bubbles in the pump tube
Put the liquid from the tube
tip to the pump tube.
Summary of Contents for ZA3000 Series
Page 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Page 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Page 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Page 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...