@
[email protected]
+91 7046223333 , 9427301436
www.utplindia.in
www
Model No. : UT-1405
TEMP. AND DIFF. PRESS. INDICATOR
[CLEAN ROOM INDICATOR]
INSTRUCTION MANUAL
M
UT-1405
888.8
5.0
2
8888
0
3
Page 1: ...sales utplindia in 91 7046223333 9427301436 www utplindia in www Model No UT 1405 TEMP AND DIFF PRESS INDICATOR CLEAN ROOM INDICATOR INSTRUCTION MANUAL M UT 1405 UT 1405 888 8 5 0 2 8888 0 3...
Page 2: ...AGE Power Supply Wiring Requirements Applying 240V AC to a controller rated for 12 24VDC will severely damage the controller and is a fire and smoke hazard This controller is suitable for connection t...
Page 3: ...CHART 4 1 CONFIGURATION MODE 4 OPERATIONAL MODES 4 2 SETUP MODE 4 3 CALIBRATION MODE 4 4 CHANGE PASSWORD MODE 4 5 BUZZER MODE 5 COMMUNICATION MODBUS RTU PROTOCOL 5 1 INTRODUCTION 5 2 MODBUS CONFIGURAT...
Page 4: ...INSTALLATION Indicator Frame NOTE All Dimensions are in mm The UNITECH s Clean Room Indicator T DP UT 1405 has been specially developed keeping process measurement requirements of Clean rooms paramet...
Page 5: ...ired in the back wall or partition 125 H X 100 W 3 Fix the frame on surface by screws 4 Insert the Power Communication Cables 100 125 100 145 155 110 100 145 155 110 W all O R Partition surface FRAM E...
Page 6: ...w scale Full range as per input selection General Specificaion Power Supply 24VDC Power Consumption 4 Watts Maximum Prog parameters Modbus RTU RS 485 Isolated Communication Baud rate Parity Node addre...
Page 7: ...meproof model available for Hazardous applications This instrument enables you to measure parameters in the following fields HVAC System Heating ventilation Air conditioning cooling Pharmaceutical ind...
Page 8: ...in Programming Mode NORMAL VIEWS ENTER KEY ENTER Key allows enter into the Parameter moving from one parameter to another and accept saving a new value of parameter or a status of parameter UP KEY UP...
Page 9: ...bUZZ 1SP1 1SP2 1SP3 1SP4 2SP1 2SP2 2SP3 2SP4 Password Protected CAL CAL1 CAL RANG WARNING INJURY Programming parameter should be accessed only by Personnel who are technically competent to do so Local...
Page 10: ...TEMP Alarm 1 Type 1sp1 Default type should be Lo Values are Low Low Low High High High LL LO H1GH HH II Input 1 TEMP Alarm 2 Type 1sp2 Values are Low Low Low High High High LL LO H1GH HH III Input 1 T...
Page 11: ...RM STATUS ENABLE DISABLE aLST Enable OR Disable Alarm of particular Input Press key to access sub parameters the sub parameters are I Input 1 TEMP Alarm 1 1sp1 Values are Enable Disable En D1s Values...
Page 12: ...ETER COMM Device Active Master Slave Communication configuration parameter Press key to access sub parameters the sub parameters are Device Address for Communication 6 INPUT DAMPING daMp Damping param...
Page 13: ...the sub parameters are III Hours 24Hr hour Values are between to 0000 0024 Real time Hours Present Calender Date IV Day Date day Values are between to 0000 0031 Present Month IV Month MOn Values are b...
Page 14: ...DP Alarm 2 Set Point 2sp2 VII Input 2 DP Alarm 3 Set Point 2sp3 VIII Input 2 DP Alarm 4 Set Point 2sp4 V Input 2 DP Alarm 1 Set Point 2sp1 2 Input 1 TEMP Alarm 1 Set Point 1sp1 Set any value between...
Page 15: ...rd for access Setup mode 1234 II Input 1 TEMP High Calibration 1p1s Apply to High Span signal to Input 1 TEMP and press key IV Input 2 DP High Calibration 1p2s Apply to High Span signal to Input 2 DP...
Page 16: ...sword 1234 Set the new Password for Buzzer Mode BUZZ 5 BUZZER MODE PASSWORD BUZZ Change Password Mode default password 1234 Set the new Password for Change Password Mode pASC 6 CHANGE PASSWORD MODE PA...
Page 17: ...405 Device 2 UT 1405 Device 3 UT 1405 Device 256 2 core 1 sq mm shielded cable COMPUTER SCADA TEMP ALARM 2 SET POINT TEMP ALARM 3 SET POINT TEMP ALARM 4 SET POINT 40007 40009 40008 40010 40011 40012 4...
Page 18: ......