8.6.1
8 - 7
(1) Confirm that the reference signal is at nearly zero level.
If the signal is deviated considerably from the zero level,
click the [Auto-Zero] button. The baseline returns to zero.
Fig. 8-4 [Auto-Zero] Button
(2) Open the lid of atomizer unit and inject a certain amount of
the sample with a micro-pipet or the like. Then close the
lid carefully.
NOTE
: Be very careful during injection as the absorbance
changes due to an error in the injection volume, etc.
(3) Click the [Start] button in the tool box.
Fig. 8-5 [Start] Button
(4) The temperature program starts to carry out measurement.
The status area in the Instrument Monitor Bar contains the
ID of a sample under measurement, the current
measurement stage and a time bar to indicate progress of
each measurement stage..
(5) At completion of all measurements, “Ready” appears on the
status area.
(6) Check the level of reference signal. If significantly
deviated from the zero level, click [Auto-Zero] button as
required.
(7) Repeat the procedure (2) onward.
8.6.1 Starting
Measurement