8.3 Operation
8 - 4
8.3 Operation
Follow the instructions below, assuming the instrument is
diagnosed on all items in the flame mode.
The lamp or standard solution may not be required if all
diagnostic tests are not performed.
(1) Install the lamps into the lamp turret.
The Hg and Cu lamps should be set in No. 8 and 7,
respectively.
(2) Use the standard burner and provide the “Air-C
2
H
2
” flame.
(1) Click the [Diagnosis] button in the toolbox to display the
Diagnosis window.
Fig. 8-1 [Diagnosis] Button
(2) Click the [Measurement Mode] button on the toolbar, or
[Measurement Mode] in the [Method Setup] menu.
The Measurement Mode dialog box opens.
In Measurement Mode, select [Flame].
Fig. 8-2 [Measurement Mode] Button
8.3.1 Preparation
8.3.2 Setting
of
Conditions