10.1 Setting of Temperature Program Test Conditions
10 - 14
REF detection start level
REF detection end level
Ashing - Background
Detection
start
Detection
end
Ashing
5s
Reference signal
Time(s)
Fig. 10-10 Background Detection in the Ash Test
NOTES
: 1. Background detection may fail because the
reference signal does not rise adequately depending
on the kind of sample, sample injection volume,
measuring wavelength, cuvette type, etc.
In this case, the same ashing time as the [Ramp
Time (s)] value set in an ashing test will be
adopted.
2. Generally, a sample containing much concomitant
matter tends to cause a rise in the reference signal
at the ashing stage.
On the other hand, the reference signal may be
too low to detect depending on a sample
containing an extremely small amount of
concomitant matter or measuring wavelength.
In such cases, the same ashing time as the [Ramp
Time (s)] value set in an ashing test will be
adopted.
3. Shortly after turning on the light source lamp, the
reference signal fluctuates due to a change in the
lamp energy.
The lamp requires a warm-up period of about 10
minutes after turning it on. Measurement should
be started after the reference signal has stabilized.
Summary of Contents for ZA3000 Series
Page 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Page 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Page 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Page 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...