11. TROUBLESHOOTING
11 - 2
(cont’d)
Symptom
Cause
Check and Remedy
Repeatability is
poor.
The nozzle of autosampler is
inappropriate.
Check if the depth of nozzle in the
graphite cuvette is optimum.
Check if the nozzle has an adequate flow
cut performance. If the performance is
poor due to contamination, wipe off
contaminants or pull out the tubing and
cut off the tip to renew it.
Air bubbles are formed in the
flow path of autosampler.
Check for air bubbles at mili/micro-
syringe. If air bubbles are formed,
discharge them by executing [Move Milli-
Syringe Up and Down] repeatedly.
Recovery is poor.
Interference by coexisting
substances in sample
Reduce coexisting substances’
concentration by dilution or extraction.
Addition of a matrix modifier may be
required depending on element.
Sample viscosity is too high.
Check the state of aspiration while
aspirating the sample and reduce its
viscosity by dilution or decomposition.
Or slow the injection speed of
autosampler.
Baseline is
unstable.
The lamp has been deteriorated.
Check the lamp spectrum and replace
the lamp when required.
Lamp current is inappropriate.
Correct the lamp current.
The lamp (element) is wrong.
Check the lamp house and replace the
lamp with a correct one.
Noise is large.
Intensity of lamp is inadequate.
Raise the lamp current of hollow cathode
lamp or replace the lamp with a new one
when it has reached the end of lifetime.
Time constant is inappropriate.
Change to an appropriate time constant.
The service life of
graphite cuvette is
short.
Water has leaked in the atomizer
unit.
Check with the electrode removed and
wipe off water completely.
Each temperature of atomizing
and cleaning stages is too high.
Check the atomizing and cleaning stages
of temperature program, and set an
appropriate heating temperature and hold
time.
Flow rate of exhaust duct is too
high.
Reduce flow rate of the duct.
Summary of Contents for ZA3000 Series
Page 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Page 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Page 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Page 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...