1.6
1 - 11
WARNING
Beware of Toxic Gas!
Inhalation of toxic gas could be injurious to human health.
Inhaling the vapor emitted from the sample may cause
inflammation of the respiratory tract.
Be sure to operate the ventilating fan to discharge the
vapor emitted from the sample through the exhaust duct.
1.6 Power-on Exhaust Duct
Power on the exhaust duct and operate the fan.
The air volume of the exhaust duct should be 10-25m
3
/hr. Adjust
the air volume by using the damper, etc. if the volume is large.
Large air volume may cause the argon gas (sheath gas) (which
is used to protect a cuvette) to be sucked in too much, resulting
in shorter service life of the cuvette.
WARNING
Beware of Toxic Gas!
Inhalation of toxic gas could be injurious to human health.
Inhaling the vapor emitted from the sample may cause
inflammation of the respiratory tract.
Be sure to operate the ventilating fan to discharge the
vapor emitted from the sample through the exhaust duct.