6.5.1
6 - 15
4. When a dilution factor is specified in one of Check
Recovery and Check Sample, the value will be
automatically assigned in the other checks.
Action:
Specify any of the processes below in case recovery
does not satisfy Recovery Tolerance.
Continue ............... Continues the sequence of
unknown sample measurement
with a message.
End ....................... Terminates measurement for
current element.
When an element to be measured
next is set, however, it will be
measured.
Dilute Sample ....... Dilutes the unknown sample and
performs recovery check again.
NOTES
: 1. When dilution has been made in Check Sample,
Check Recovery is performed with the diluted
unknown sample alone.
2. If recovery for diluted sample does not satisfy
Recovery Tolerance again, sample dilution is not
repeated and the next sample measurement will
be performed.
(3) In the window appearing when clicking the [Modifier/QC
Cup No.] tab in the Autosampler window, enter the cup No.
of Check Recovery (spike sample).
The cup number specified for Check Recovery is indicated
as “R” in the window appearing when clicking the [Sample
Map] tab.