1 - 12
1.5.4 Installation of Atomizer
Attach heat shield plates to the magnet.
DANGER
This instrument utilizes a 0.9 tesla permanent
magnet in the burner unit. There is a danger of
faulty operation of a pacemaker due to the strong
magnetic field produced. A person who uses a
pacemaker should not come within 1 meter of the
instrument, or should avoid handling the instrument
altogether.
CAUTION
The strong magnet field generated by the flame
magnet can cause severe injury. Do not bring metal
objects including iron, such as a screwdriver, near
the magnet.
NOTICE
The strong magnet field generated by the flame
magnet may cause damage to a watch or magnetic
card and so on. Avoid bringing such objects near
the magnet.
(1) Turn the burner height adjustment knob of AAS main unit
counter-clockwise until the burner height reaches the lowest
level.
Fig. 1-12 Lowering Burner Position
Burner height
adjustment knob
Summary of Contents for ZA3000 Series
Page 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Page 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Page 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Page 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...