3.4 Wavelength Adjustment
3 - 54
Condition setting requires about 1 to 3 minutes.
When [Ready] appears in the status field, this step is
completed.
(4) On the menu bar, open the [Instrument Control] menu and
click [Lamp Spectrum]. Make sure the peak at the
measuring wavelength appears at the center of the
displayed spectral graph, and then click the [OK] button.
Line spectra obtained in “Next Data” or “Previous Data” can
be checked. The displayed line spectrum can be printed
with the printer by clicking [Print].
Fig. 3-42 Lamp Spectrum Dialog Box
CAUTION
Beware of Intense Light!
The hollow cathode lamp emits intense ultraviolet rays.
Directly gazing at the lamp could therefore result in
impaired vision. Do not look directly at the light from the
hollow cathode lamp.