4.5.11
4 - 25
(1) In Data Table, select the row of the sample to be specified
and click [Sample Blank] - [Set Sample Blank] in the [Data]
menu.
(2) A dialog box for confirmation opens. Click the
[Recalculate] button.
(3) To the sample ID set as a sample blank, “S” is attached (for
example, UNK-001S), but the analytical values of that
sample remain unchanged. For the samples after the
sample blank, analytical values after subtracting those of
the sample blank will be indicated.
NOTE
: STD sample cannot be specified as a sample blank.
(4) For canceling the sample blank specification, select the row
of the corresponding sample and click [Sample Blank] -
[Cancel Sample Blank] in the [Data] menu.
(5) A dialog box for confirmation opens. Click the
[Recalculate] button.