1 - 6
1.5.2 Installation of Separator
(1) Loosen the screws on the left side face of HFS-4, fit the
separator fixture (part No. 7J1-7185) and tighten the screws
again.
Fig. 1-4 Mounting of Separator Fixture
CAUTION
The separator is made of glass.
If the separator is broken, you may suffer an injury.
Adequate attention should be paid when handling
the separator.
(2) Connect the drain tubes (part No. 171-6327) to the separator
(part No. 7J1-7400) provided as an accessory.
(3) Put the pure water in the separator.
(4) Insert the fluoro-rubber plug coming from the left side face of
HFS-4 into the separator by about 13 mm. Then, attach the
separator to the separator fixture.