10.1
10 - 11
Set the following items.
Start/End Temp. (°C)
Enter the start/end temperature of ashing test.
Usually enter a value within about 400 to 500
C.
Input range: 50 to 2800
C
Max. Temp. (
C)
Enter the maximum temperature when raising the
ashing temperature at the set intervals in the [Ash Test].
Usually enter a value within about 1000 to 1600°C.
Input range: 50 to 2800°C
Interval (
C)
Enter a heating interval for raising the start/end
temperature at the set intervals in the [Ash Test].
Usually enter a value of about 100 or 200
C.
Input range: 50 to 500
C
NOTES
: 1. If the difference between the start/end temperature
and maximum ashing temperature is not adequate
in comparison with this interval, an error will be
displayed.
Correct your entries so that (maximum ashing
temperature - start/end temperature) is larger than
or equal to this interval.
2. If the difference obtained by subtracting the
start/end temperature from the maximum ashing
temperature cannot be divided by the interval
value, the program will be executed as shown
below.
Example Execution of Ash Test
Ash Test Conditions: [Start/End Temp.],
[Max. Temp.] and
[Interval] are set at 400°C,
700°C and 200°C.