1.4.2
1 - 7
DANGER
Malfunction of Pacemaker!
There is a danger of faulty operation of a pacemaker due
to the strong magnetic field produced.
A person who uses a pacemaker should not come within 1
meter of the instrument, or should avoid handling the
instrument altogether.
CAUTION
Ignition of the Graphite Furnace!
There is a danger of ignition and resulting burns due to
residual moisture.
If moisture remains in the graphite furnace after cleaning
the interior, the moisture may be decomposed and
produce a hydrogen flame when heating the unit. So if
water is spilled or cooling water leaks, be sure to wipe it up
thoroughly.
CAUTION
Strong Magnetic Fields!
Be careful not to let your hand get attracted to the magnet.
Do not bring metal objects including iron, such as a
screwdriver, near the magnet.
CAUTION
Strong Magnetic Fields!
The graphite furnace utilizes a magnet.
The strong magnetic field produced may cause damage to
a watch, magnetic card and so on.
Avoid bringing such objects near the magnet.
Mount the selected graphite cuvette to the atomizer unit in the
following procedure. This instrument is installed with LED lights
which illuminate the atomized part to improve the visibility. Do
not interrupt the light by hands.
(1) Open the lid of atomizer unit, holding its white resin part.
1.4.2 Mounting
of
Graphite Cuvette
Summary of Contents for ZA3000 Series
Page 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Page 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Page 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Page 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...