1.4 Graphite Cuvette Installation
1 - 6
NOTES
: 1. The graphite cuvette is worn and deteriorated due
to sublimation of graphite or its reaction with the
sample vapor, etc. through repeated
measurements. If the graphite cuvette has
significantly deteriorated, atomic absorption
sensitivity fluctuates and reproducibility is
degraded. In such a case, the graphite cuvette
should be replaced with a new one.
The cumulative number of graphite furnace
operations is displayed on the monitor window.
Before start of measurement, check the displayed
number and replace the graphite cuvette with a
new one when necessary.
Note that attention should be paid to the progress
of deterioration of the graphite cuvette since its
lifetime varies with sample constituents.
In analyses of samples similar to an aqueous
solution, the Pyro Tube HR has an approximate
lifetime of 1,000 measurements.
2. The heating temperature varies depending on the
resistance value of each cuvette. Do not perform
measurement by using any cuvettes other than
those configured by this instrument.