3 - 10
(2) Check that the separator contains pure water.
Fig. 3-8 Checking Separator
(3) Connect Ar gas to the HFS-4 and supply the gas.
Turn on the power switch of the HFS-4 and confirm that the
power lamp lights up and the pump starts to rotate.
NOTE:
If the argon gas is not supplied to the HFS-4, the system
does not run. Confirm the supply of argon when the pump
does not start to rotate.
(4) Visually check to ensure that liquid is flowing in each pump tube.
(5) Ignite the flame.
NOTE 1:
Check the gas leak before use, and power-on of exhaust
duct.
NOTE 2:
The fuel flow rate should to be set at about 1.2 L/min
when the absorption cell is used. A flow rate of normal
flame condition (about 2.0 L/min) may result in ignition
failure.
NOTE 3:
Although the color of the thermal tape on the flame
magnet may be change to black, that is not abnormal.
NOTE 4:
Do not extinguish the flame without disconnecting the
Silicone tube from the separator. Otherwise the cell may
crack.