3.1 Generation of Measurement Conditions
3 - 16
Fig. 3-9 Pasting the dry stage (2)
NOTE
: When [Paste Dry Stage] is executed, all of the Dry
Stage is overwritten by the setting of the copied Dry
Stage (Stage No.1 to 4).
Temperature program development
This is a function to determine a temperature program
that can be obtained from the maximum absorbance or
the minimum RSD value by performing automatic
measurement while changing the drying, ashing and
atomizing temperatures stepwise. Since multiple
temperature program patterns can be automatically
measured sequentially, this function can be used to
examine a temperature program to determine analytical
conditions. For details, refer to 10 “Temperature
Program Development Function.”
Bumping Detection of the sample
In the dry stage, the bumping of the sample due to rapid
heating is detected from the obtained signal data when
the sample in the cuvette is heated. The alert mark “P”
is given to the measurement results of the sample for
which bumping was detected. To use this function,
check the checkbox (
).For details, refer to 5.7
“Sample Bumping Detection” in Chapter 5.