7.3 Closing AAS program
7 - 4
Fig. 7-5 Dialog Box for Acetylene Gas
NOTE
: Many of the samples include acid. Leaving them may
cause the instrument to rust.
(4) The message of an acetylene gas main cylinder closed
check will appear. Close the main valve of the acetylene
gas before working on the next task.
Fig. 7-6 Dialog Box for Acetylene Gas
(5) A message to terminate the atomic absorption
spectrophotometer appears. Click [Yes] to close.
Fig. 7-7 Dialog Box for Terminating Software
(6) If any measurement result data is not yet saved, a
confirmation message for saving will be displayed.
Select a data saving option.
NOTE
: If the [No] button is clicked in answer to the data
saving confirmation message, the measurement result
data will be lost (not saved).