10.3
10 - 19
10.3 Heating Temperature Graph
Temp. Program Development function displays the heating
temperature graph on the location of the working curve graph on
the [Data Collection Monitor] and [Data Process] windows.
Fig. 10-14
Heating Temperature Graph
The graph switching button is available on the upper part of the
heating temperature graph.
Drying temperature graph button
Drying temperature graph is displayed. Reproducibility of the
drying starting temperature is displayed.
Ashing temperature graph button
Ashing temperature graph is displayed. Absorption sensitivity
in the ashing temperature is displayed.
Atomization temperature (ABS) graph button
Atomization temperature (ABS) graph button is displayed.
Absorption sensitivity in the atomization temperature is
displayed.
Atomization temperature (RSD) graph button
Atomization temperature (RSD) graph button is displayed.
Reproducibility in the atomization temperature is displayed.
Click each button of each measurement item for which the Temp.
Program Development finished to display each heating
temperature graph.