3.1.5
3 - 13
Gas
Type
Specifies the kind of gas.
Specify Normal (argon gas) usually.
Change to the alternate gas (refer to 1.3 “Supply of
Gasses”) when required for speeding up oxidation of
organic substances in a sample (optional alternate gas
unit is necessary). If the alternate gas is used, set the
temperature below 500°C.
A dry and/or ash stage can be added in the following
procedure.
Click the number of the target stage so that the entire row
of the number turns blue. Then, click the [Recall] button.
For deleting the set stage, also click the number of the
target stage so that the entire row of the number turns blue,
and then click the [Delete] button.
(4) Other
functions
Copy Dry Stage
Setting of the Dry Stage of selected analyte element can
be copied. Since the dry stage is determined by the
properties of the sample liquid, the same conditions can
be used for different analyte elements if the samples are
the same.
Click the [Copy Dry Stage] button when you paste the
indicated Dry Stage on another analyte element.