6.3.2
6 - 9
2. If working curve is regenerated during
measurement in the simple STD addition method,
data of the first unknown sample can be
calculated again.
However, the value calculated first will be reported
for the first unknown sample and the recalculated
curve will be used for subsequent samples only.
3. QC sample measurement is performed after
measurement of a group of unknown samples.
Therefore, if the [Next Element] button or the
[Close Sequence] button is clicked before
measurement of the last QC sample, a prompting
message will appear asking for execution of QC
sample measurement.
The following messages are printed in the report:
QC measurement value exceeds the permissible range.
Measurement is finished.
Measure all standard samples anew and prepare working
curves again. Re-measure sample groups involved in error.
6.3.2 Contents to be
printed on the
report
Summary of Contents for ZA3000 Series
Page 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Page 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Page 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Page 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...