3.6 Flowing of Cooling Water
Supply cooling water to the instrument. When the cooling water
is supplied for the first time on the day of measurement, the
following dialog box will appear. Check for water leaks. If no
leaks are found, click [OK].
Fig. 3-45 Dialog Box for Checking for
Water Leaks from the Sample Line
NOTES
: 1. A water flow of about 2 L/min is required within a
water pressure range from 35 to 50 kPa (about
0.35 to 0.5 kgf/cm
2
). If the temperature of
cooling water is much lower than the room
temperature, condensation of water vapor in the
air may occur on the quartz window of atomizer,
resulting in degradation of the components and
defective data. Water temperature should be
limited within ±5 °C of the room temperature.
2. Cooling water should be supplied shortly before
measurement for preventing condensation.