10.2 Start of Measurement
10 - 18
10.2 Start of Measurement
(1) Specify, on the [Autosampler] window, the position of the
sample cup (cup No.) which is equivalent to the unknown
sample ID for evaluation.
Temp. Program Development function measures only one
sample specified for evaluation.
(2) Determine conditions and start measurement.
Temp. Program Development function displays a message
on the measurement replicate after clicking [Verify] button.
Fig. 10-13 Measurement Replicate
Announcement Window
NOTE
: Choose samples to be used for this function with
absorbance in the range of 0.01-0.25 and background
absorbance (REF) of 1.0 or less. An error message
“The sample is inadequate for the test” appears when
the value measured exceeds this range. In such an
event, prepare again a sample with the above
absorbance value.