9.3 Operation
9 - 6
(5) In Diagnosis Threshold, the recommended values are listed.
These values are changeable, if required, except for
[Wavelength Accuracy].
Wavelength
Accuracy
0.45 nm. No change is allowed.
Baseline Stability (Noise)
Input a permissible noise value in the change of
absorbance with a time period of 5 minutes.
0.003 to 0.02 Abs
Baseline Stability (Drift)
Input a permissible drift value of absorbance in 5
minutes.
0.002 to 0.02 Abs
Sensitivity
Input a permissible absorbance value in
measurement of Cu 20-
g/L standard solution (3
repetitive measurements).
0.01 to 0.1 Abs
Repeatability
Input a permissible relative standard deviation value
in measurement of Cu 20-
g/L standard solution (10
repetitive measurements).
RSD 3 to 30%
(6) Click the [Next] button to open the [Report Format] dialog
box.