3.9 Start of Measurement
3 - 64
3.9 Start of Measurement
DANGER
Malfunction of Pacemaker!
There is a danger of faulty operation of a pacemaker due
to the strong magnetic field produced.
A person who uses a pacemaker should not come within 1
meter of the instrument, or should avoid handling the
instrument altogether.
WARNING
Beware of Toxic Gas!
Inhalation of toxic gas could be injurious to human health.
Inhaling the vapor emitted from the sample may cause
inflammation of the respiratory tract.
Be sure to operate the ventilating fan to discharge the
vapor emitted from the sample through the exhaust duct.
CAUTION
Ignition of the Graphite Furnace!
There is a danger of ignition and resulting burns due to
residual moisture.
If moisture remains in the graphite furnace after
cleaning the interior, the moisture may be decomposed
and produce a hydrogen flame when heating the unit.
So if water is spilled or cooling water leaks, be sure to
wipe it up thoroughly.
If the cover of the graphite furnace is left open when
heating, there is a danger of ignition resulting in burns.
Be sure to close the cover of the unit when heating and
performing measurement.
CAUTION
Strong Magnetic Fields!
Be careful not to let your hand get attracted to the magnet.
Do not bring metal objects including iron, such as a
screwdriver, near the magnet.
CAUTION
Strong Magnetic Fields!
The graphite furnace utilizes a magnet.
The strong magnetic field produced may cause damage to
a watch, magnetic card and so on.
Avoid bringing such objects near the magnet.
Summary of Contents for ZA3000 Series
Page 423: ...9 3 5 9 11 Fig 9 14 Diagnosis Log Dialog Box ...
Page 449: ...10 5 Reflecting on Analysis Conditions 10 24 Fig 10 16 Method Window Unmark the checkbox ...
Page 455: ...INDEX 4 V Verify 3 34 W Working curve graph 4 1 Wavelength accuracy 9 1 ...
Page 456: ...INSTRUCTION MANUAL 7J1 9027 000 Ver 1 February 2016 MODEL HFS 4 HYDRIDE FORMATION SYSTEM ...