Measurements
R&S
®
FSW
303
User Manual 1173.9411.02 ─ 19
Note:
The EMI-specific filter types are available if the EMI (R&S
FSW-K54) measure-
ment option is installed, even if EMI measurement is not active. For details see
ter 5.13.3.1, "Resolution Bandwidth and Filter Types"
The RBW filter configured in the bandwidth settings is identical to the filter configured
in the EMI configuration.
Remote command:
RBW
Defines the resolution bandwidth. The available resolution bandwidths are specified in
the data sheet. Numeric input is always rounded to the nearest possible bandwidth.
If "AUTO" is selected, the resolution bandwidth is coupled to the selected span (for
span > 0). If the span is changed, the resolution bandwidth is automatically adjusted.
If the resolution bandwidth is defined manually, a green bullet is displayed next to the
"RBW" display in the channel bar.
For more information see
chapter 6.5.1.1, "Separating Signals by Selecting an Appro-
For measurements on I/Q data in the frequency domain, the maximum RBW is 1
MHz.
For EMI measurements using the quasipeak detector, the 1
MHz RBW filter is not
available (see
chapter 5.13.3.1, "Resolution Bandwidth and Filter Types"
Remote command:
Automatic Peak Search
If activated, a peak search is performed automatically for all active markers after each
sweep.
If Auto peak search
and
limit lines are active, the active markers are set to the peak
delta
values between the measured signal and the limit lines.
Note:
are not available for
EMI measurements.
Remote command:
Dwell Time
Sets the dwell time for the EMI marker measurement.
For more information see
chapter 5.13.3.2, "Detectors and Dwell Time"
Remote command:
Frequency Axis Scaling
Switches between linear and logarithmic scaling for the frequency axis.
Electromagnetic Interference (EMI) Measurement (R&S
FSW-K54)