Remote Commands
R&S
®
FSW
984
User Manual 1173.9411.02 ─ 19
Parameters:
<Distance>
Defines the distance that the limit line moves.
The unit depends on
Usage:
Event
Manual operation:
See
CALCulate:LIMit<k>:UPPer:SPACing
<InterpolType>
This command selects linear or logarithmic interpolation for the calculation of an upper
limit line from one horizontal point to the next.
Parameters:
<InterpolType>
LINear | LOGarithmic
*RST:
LIN
Manual operation:
See
See
CALCulate:LIMit<k>:UPPer:STATe
<State>
This command turns an upper limit line on and off.
Before you can use the command, you have to select a limit line with
Parameters:
<State>
ON | OFF
*RST:
OFF
Usage:
SCPI confirmed
Manual operation:
See
CALCulate:LIMit<k>:UPPer:THReshold
<Limit>
This command defines an absolute limit for limit lines with a relative scale.
The R&S
FSW uses the threshold for the limit check, if the limit line violates the thresh-
old.
Parameters:
<Limit>
Numeric value.
The unit depends on
on page 982.
*RST:
-200
Default unit: dBm
Manual operation:
See
Analyzing Measurements (Basics)