Search Mode for Next Peak in X Direction
..................................................................502
Search Mode for Next Peak in Y Direction
..................................................................502
.......................................................................................505
........................................................................... 505
..........................................................................................505
........................................................................................ 505
Deactivating All Search Limits
...................................................................... 506
Search Mode for Next Peak in X Direction
Selects the search mode for the next peak search within the currently selected frame.
"Left"
Determines the next maximum/minimum to the left of the current
peak.
"Absolute"
Determines the next maximum/minimum to either side of the current
peak.
"Right"
Determines the next maximum/minimum to the right of the current
peak.
Remote command:
Search Mode for Next Peak in Y Direction
Selects the search mode for the next peak search within all frames at the current
marker position.
"Up"
Determines the next maximum/minimum above the current peak (in
more recent frames).
"Absolute"
Determines the next maximum/minimum above or below the current
peak (in all frames).