
Common Measurement Settings
R&S
®
FSW
377
User Manual 1173.9411.02 ─ 19
●
"To connect a three-port mixer"
●
●
"To activate and configure the external mixer"
●
"To define a new conversion loss table"
●
"To shift the conversion loss values"
To connect a three-port mixer
External mixers can be connected at the LO OUT/IF IN and IF IN female connectors
(option R&S
FSW-B21). Both two-port and three-port mixers can be used. Connect the
mixer as follows:
Use the supplied coaxial cable to feed in the LO signal. If no external mixers are con-
nected to the R&S
FSW, cover the two front connectors LO OUT / IF IN and IF IN with
the SMA caps supplied.
1. Connect the LO OUT / IF IN output of the R&S
FSW to the LO port of the external
mixer.
2. Connect the IF IN input of the R&S
FSW to the IF port of the external mixer.
3. Feed the signal to be measured to the RF input of the external mixer.
Data Input and Output