Measurements
R&S
®
FSW
217
User Manual 1173.9411.02 ─ 19
............................................................................................... 217
..............................................................................................217
................................................................................................ 217
............................................................................................217
Power Reference Type
Defines how the reference power is calculated.
"Channel Power"
Measures the channel power within the reference range using the
integration bandwidth method (see also
Additional settings can be configured for this method.
"Peak Power"
Determines the peak power within the reference range.
Remote command:
Channel Power Settings
If the Power Reference Type "Channel Power" was selected, additional parameters
can be configured.
Tx Bandwidth ← Channel Power Settings
Defines the bandwidth used for measuring the channel power, with:
minimum span ≤ Tx Bandwidth ≤ span of reference range
Remote command:
RRC Filter State ← Channel Power Settings
Activates or deactivates the use of an RRC filter.
Remote command:
Spectrum Emission Mask (SEM) Measurement