Measurements
R&S
®
FSW
309
User Manual 1173.9411.02 ─ 19
2. Define the frequency range of the measurement.
a) Press the FREQ key.
b) Press the "Start Frequency" softkey and enter a frequency of 150
kHz.
c) Press the "Stop Frequency" softkey and enter a frequency of 1
GHz.
The R&S
FSW scales the horizontal axis accordingly.
3. Press the MEAS key on the front panel and select the "EMI" measurement.
The EMI main menu is displayed.
4. Select the "EMI Config" softkey and define the resolution bandwidth and filter type
to be used for the measurement.
By default, the R&S
FSW uses a filter with a 3
db bandwidth. EMI measurements
usually require a filter with a 6R&S
FSWdB bandwidth.
5. Define the dwell time for which each marker position is measured during the final
measurement.
6. To obtain an overview of exceptional values in the input signal during the initial
measurement, activate the "Auto Peak Search".
7. Select the measurement bandwidth.
a) Select the "Res BW CISPR" softkey.
A CISPR (6
dB) filter is configured.
b) Set the bandwidth to 1
MHz.
The R&S
FSW shows the currently selected resolution bandwidth in the diagram
header.
8. Configure the traces for the initial EMI measurement.
a) Press the TRACE key.
b) Press the "Trace Config" softkey to configure two traces.
c) Define the detectors to use for the initial measurement. Select the peak detec-
tor for trace 1 and the average detector for trace 2.
The peak detector ensures that the detected peak levels in the frequency range
covered by one pixel are displayed.
The R&S
FSW now displays two traces. Trace 1 shows the peak values, trace 2
shows the average values.
9. Increase the number of sweep points for the EMI measurement.
a) Press the SWEEP key on the front panel.
b) Select the "Sweep Config" softkey.
c) Set the number of "Sweep Points" to
200000
.
10. Press the AMPT key, then select the "Amplitude Config" softkey and, in the "Ampli-
tude" dialog box, select
V
as the "Unit".
Performing the measurement
1. Configure the EMI measurement markers. In this example we will use 6 markers.
Electromagnetic Interference (EMI) Measurement (R&S
FSW-K54)