
Common Analysis and Display Functions
R&S
®
FSW
496
User Manual 1173.9411.02 ─ 19
Assigning the Marker to a Trace
The "Trace" setting assigns the selected marker to an active trace. The trace deter-
mines which value the marker shows at the marker position. If the marker was previ-
ously assigned to a different trace, the marker remains on the previous frequency or
time, but indicates the value of the new trace.
The marker can also be assigned to the currently active trace using the "Marker to
Trace" softkey in the "Marker" menu.
If a trace is turned off, the assigned markers and marker functions are also deactiva-
ted.
Remote command:
Select Marker
Opens a dialog box to select and activate or deactivate one or more markers quickly.
Remote command:
Marker selected via suffix <m> in remote commands.
All Markers Off
Deactivates all markers in one step.
Remote command:
General Marker Settings
Some general marker settings allow you to influence the marker behavior for all mark-
ers.
These settings are located in the "Marker Settings" tab of the "Marker" dialog box. To
display this tab, do one of the following:
●
Press the MKR key, then select the "Marker Config" softkey.
●
In the "Overview", select "Analysis", and switch to the vertical "Marker" tab. Then
select the horizontal "Marker Settings" tab.
Marker Usage