Common Measurement Settings
R&S
®
FSW
353
User Manual 1173.9411.02 ─ 19
2. Connect the signal generator output to the RF INPUT connector on the front panel
of the R&S
FSW.
3. Adapt the measurement range of the R&S
FSW to the filter to be tested. In this
measurement, define the following settings:
a) Press the FREQ key, select "Frequency Config" and enter "Frequency Start":
100
MHz
.
b) Enter "Frequency Stop":
300
MHz
4. Press the INPUT/OUTPUT key and select "External Generator Config".
5. In the "Interface Configuration" sub-tab, select "Generator Type":"SMU06".
6. Select "Reference: External" to synchronize the analyzer with the generator.
7. Switch to the "Measurement Configuration" sub-tab.
8. Set the "Source State" to "On".
9. Define the generator output level as the "Source Power":
-20
dBm
.
10. Set the "Coupling State" to "Auto".
The "Result Frequency Start" value for the generator is indicated as
100.0
MHz
.
The "Result Frequency Stop" value is indicated as
300.0
MHz
.
11. Switch to the "Source Calibration" sub-tab.
12. Select the "Source Calibration Type": "Transmission" to perform a calibration
sweep and store a reference trace for the measurement setup.
Fig. 6-7: Measurement results from generator, analyzer and connecting cables
13. Select "Source Calibration Normalize": "On" to set the measurement results for the
current setup to 0, thus eliminating all effects from the generator, the analyzer and
Data Input and Output