Remote Commands
R&S
®
FSW
976
User Manual 1173.9411.02 ─ 19
Parameters:
<Position>
The value range is variable.
You can use any unit you want, the R&S
FSW then converts the
unit to the currently selected unit. If you omit a unit, the
R&S
FSW uses the currently selected unit.
*RST:
(state is OFF)
Example:
CALC:DLIN -20dBm
Positions the display line at -20 dBm.
Manual operation:
See
CALCulate<n>:DLINe<k>:STATe
<State>
This command turns a display line on and off
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:DLIN2:STAT ON
Turns on display line 2.
CALCulate<n>:FLINe<k>
<Frequency>
This command defines the position of a frequency line.
Parameters:
<Frequency>
Note that you can not set a frequency line to a position that is
outside the current span.
Range:
0 Hz to Fmax
*RST:
(STATe to OFF)
Example:
CALC:FLIN2 120MHz
Sets frequency line 2 to a frequency of 120 MHz.
Manual operation:
See
CALCulate<n>:FLINe<k>:STATe
<State>
This command turns a frequency line on and off
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:FLIN2:STAT ON
Turns frequency line 2 on.
Analyzing Measurements (Basics)