Network and Remote Operation
R&S
®
FSW
657
User Manual 1173.9411.02 ─ 19
LXI Manufacturer Description
Instrument description of the R&S
FSW
Remote command:
LAN Reset
Resets the LAN configuration to its default settings (LCI function).
According to the LXI standard, an LCI must set the following parameters to a default
state.
Parameter
Value
TCP/IP Mode
DHCP + Auto IP Address
Dynamic DNS
Enabled
ICMP Ping
Enabled
Password for LAN configuration
LxiWebIfc
The LAN settings are configured in the "Network" tab of the "N Remote" dialog
box or using the instrument's LXI Browser interface.
Remote command:
10.5 How to Set Up a Network and Remote Control
Risk of network failure
Before connecting the instrument to the network or configuring the network, consult
your network administrator. Errors may affect the entire network.
Remote operation
Remote operation of the instrument from a connected computer is possible using SCPI
commands (see
chapter 10.1.2, "SCPI (Standard Commands for Programmable Instru-
on page 614). Sending remote commands requires the instrument to be con-
figured in a LAN network or connected to a PC via the GPIB interface as described in
chapter 10.5.1, "How to Configure a Network"
Remote Desktop
In production test and measurement, a common requirement is central monitoring of
the T&M instruments for remote maintenance and remote diagnostics. Equipped with
the Remote Desktop software of Windows, the R&S
FSW ideally meets requirements
How to Set Up a Network and Remote Control