![Rohde & Schwarz FSW13 Скачать руководство пользователя страница 849](http://html.mh-extra.com/html/rohde-and-schwarz/fsw13/fsw13_user-manual_1477847849.webp)
Remote Commands
R&S
®
FSW
849
User Manual 1173.9411.02 ─ 19
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]
<Range>
This command defines the display range of the y-axis.
Note that the command works only for a logarithmic scaling. You can select the scaling
with
.
Parameters:
<Range>
Range:
1 dB to 200 dB
*RST:
100 dB
Example:
DISP:TRAC:Y 110dB
Usage:
SCPI confirmed
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:AUTO ONCE
Automatic scaling of the y-axis is performed once, then switched off again.
Usage:
SCPI confirmed
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:MODE
<Mode>
This command selects the type of scaling of the y-axis.
When the display update during remote control is off, this command has no immediate
effect.
Parameters:
<Mode>
ABSolute
absolute scaling of the y-axis
RELative
relative scaling of the y-axis
*RST:
ABSolute
Example:
DISP:TRAC:Y:MODE REL
Manual operation:
See
DISPlay[:WINDow<n>]:TRACe:Y[:SCALe]:RPOSition
<Position>
This command defines the vertical position of the reference level on the display grid.
The R&S
FSW adjusts the scaling of the y-axis accordingly.
For measurements with the external generator (R&S
FSW-B10) the command defines
the position of the reference value.