Common Analysis and Display Functions
R&S
®
FSW
533
User Manual 1173.9411.02 ─ 19
Margin
Defines a margin for the limit line. The default setting is 0 dB (i.e. no margin).
chapter 7.5.2, "Basics on Limit Lines"
Remote command:
X-Axis
Describes the horizontal axis on which the data points of the limit line are defined.
Includes the following settings:
●
Domain:
–
"Hz": for frequency domain
–
"s": for time domain
●
Scaling mode: absolute or relative (Hz/s/%) values
For relative values, the frequencies are referred to the currently set center fre-
quency. In the zero span mode, the left boundary of the diagram is used as the ref-
erence.
●
Scaling: linear or logarithmic
Remote command:
Y-Axis
Describes the vertical axis on which the data points of the limit line are defined.
Includes the following settings:
●
Level unit
●
Scaling mode: absolute or relative (dB/%) values
Relative limit values refer to the reference level.
●
Limit type: upper or lower limit; values must stay above the lower limit and below
the upper limit to pass the limit check
Remote command:
Data points
Each limit line is defined by a minimum of 2 and a maximum of 200 data points. Each
data point is defined by its position (x-axis) and value (y-value). Data points must be
defined in ascending order. The same position can have two different values.
Remote command:
Display and Limit Lines