Remote Commands
R&S
®
FSW
982
User Manual 1173.9411.02 ─ 19
The R&S
FSW uses the threshold for the limit check, if the limit line violates the thresh-
old.
Parameters:
<Threshold>
Numeric value.
The unit depends on
on page 982.
*RST:
-200 dBm
Manual operation:
See
CALCulate:LIMit<k>:NAME
<Name>
This command selects a limit line that already exists or defines a name for a new limit
line.
Parameters:
<Name>
String containing the limit line name.
*RST:
REM1 to REM8 for lines 1 to 8
Manual operation:
See
CALCulate:LIMit<k>:UNIT
<Unit>
This command defines the unit of a limit line.
Parameters:
<Unit>
DBM | DBPW | WATT | DBUV | DBMV | VOLT | DBUA |
AMPere | DB | DBUV_M | DBUA_M | (unitless)
If you select dB as the limit line unit, the command automatically
turns the limit line into a relative limit line.
*RST:
DBM
Manual operation:
See
CALCulate:LIMit<k>:UPPer[:DATA]
<LimitLinePoints>
This command defines the vertical definition points of an upper limit line.
Parameters:
<LimitLinePoints>
Variable number of level values.
Note that the number of vertical values has to be the same as
the number of horizontal values set with
FSW either adds
missing values or ignores surplus values.
The unit depends on
on page 982.
*RST:
Limit line state is OFF
Usage:
SCPI confirmed
Analyzing Measurements (Basics)