Remote Commands
R&S
®
FSW
965
User Manual 1173.9411.02 ─ 19
11.8.3.14
Marker Demodulation
The following commands control the demodulation of AM and FM signals at the marker
position.
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:CONTinuous
................................... 965
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:HOLDoff
..........................................965
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:SELect
............................................965
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation[:STATe]
.......................................... 966
.....................................................................................966
................................................................................. 966
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:CONTinuous
<State>
This command turns continuous demodulation of the signal at the marker position in
the frequency domain on and off.
In the time domain continuous demodulation is always on.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC2:MARK3:FUNC:DEM:CONT ON
Switches on the continuous demodulation.
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:HOLDoff
<Duration>
This command defines for how long the the signal at the marker position is demodula-
ted.
In the time domain continuous demodulation is always on.
Parameters:
<Duration>
Range:
10 ms to 1000 s
*RST:
Marker demodulation = OFF
Example:
CALC:MARK:FUNC:DEM:HOLD 3s
Manual operation:
See
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:SELect
<DemodMode>
This command selects the demodulation mode at the marker position.
Parameters:
<DemodMode>
AM
AM demodulation
FM
FM demodulation
*RST:
AM
Analyzing Measurements (Basics)