Measurements
R&S
®
FSW
113
User Manual 1173.9411.02 ─ 19
OBW
Measures the occupied bandwidth, i.e. the bandwidth which must contain a defined
percentage of the power, and opens a submenu to configure the measurement. For
details see
chapter 5.5, "Occupied Bandwidth Measurement (OBW)"
OBW measurement is only possible in the frequency domain (span > 0).
Remote command:
CALC:MARK:FUNC:POW:SEL OBW
Results:
CALC:MARK:FUNC:POW:RES? OBW
, see
chapter 11.5.5, "Measuring the Occupied Bandwidth"
Spectrum Emission Mask
Activates a Spectrum Emission Mask (SEM) measurement, which monitors compliance
with a spectral mask, and opens a submenu to configure the measurement.
For details see
chapter 5.6, "Spectrum Emission Mask (SEM) Measurement"
Remote command:
SENS:SWE:MODE ESP
, see
Results:
CALC:MARK:FUNC:POW:RES? CPOW | PPOW
, see
CALC:LIM:FAIL?
, see
chapter 11.5.6, "Measuring the Spectrum Emission Mask"
Spurious Emissions
Activates the Spurious Emissions measurement, which monitors unwanted RF prod-
ucts outside the assigned frequency band generated by an amplifier. A submenu to
configure the measurement is opened.
For details see
chapter 5.7, "Spurious Emissions Measurement"
Remote command:
SENS:SWE:MODE LIST
, see
Results:
TRAC:DATA? SPUR
, see
chapter 11.5.7, "Measuring Spurious Emissions"
APD
Measures the amplitude probability density (APD) and opens a submenu to configure
the measurement.
Available Measurement Functions