Common Analysis and Display Functions
R&S
®
FSW
519
User Manual 1173.9411.02 ─ 19
For more information see
chapter 7.4.1.4, "Searching for Signal Peaks"
Remote command:
Displaying Marker Numbers
By default, the marker numbers are indicated in the diagram so you can find the peaks
from the list. However, for large numbers of peaks the marker numbers may decrease
readability; in this case, deactivate the marker number display.
Remote command:
Exporting the Peak List
The peak list can be exported to an ASCII file (
.DAT
) for analysis in an external appli-
cation.
Remote command:
7.4.3.9
Deactivating All Marker Functions
All special marker functions can be deactivated in one step.
Use the "All Functions Off" button in the "Marker Functions" dialog box.
7.4.4 How to Work With Markers
The following step-by-step instructions demonstrate in detail how to work with markers.
How to Analyze a Signal Point in Detail
................................................................519
How to Use a Fixed Reference Marker
.................................................................520
How to Output the Demodulated Signal Accoustically
..........................................521
7.4.4.1
How to Analyze a Signal Point in Detail
Step-by-step instructions on working with markers are provided here. For details on
individual functions and settings see
chapter 7.4.2.1, "Marker Settings"
The remote commands required to perform these tasks are described in
ter 11.8.3, "Working with Markers"
When you need to analyze a characteristic point in the signal in more detail, the follow-
ing procedure can be helpful:
1. Perform a peak search to determine the characteristic point roughly by pressing
the PEAK SEARCH key.
Marker Usage