Remote Commands
R&S
®
FSW
922
User Manual 1173.9411.02 ─ 19
Parameters:
<Format>
ASCii
ASCii format, separated by commas.
This format is almost always suitable, regardless of the actual
data format. However, the data is not as compact as other for-
mats may be.
REAL,32
32-bit IEEE 754 floating-point numbers in the "definite length
block format".
In the Spectrum application, the format setting
REAL
is used for
the binary transmission of trace data.
For I/Q data, 8 bytes per sample are returned for this format set-
ting.
*RST:
ASCII
Example:
FORM REAL,32
Usage:
SCPI confirmed
TRACe<n>[:DATA]
<Trace>,<Data> | <ResultType>
This command queries current trace data and measurement results.
If you use it as a setting command, it transfers trace data from an external source to
the R&S
FSW.
The data format depends on
Parameters:
<Trace>
Selects the trace to write the data to.
TRACE1 | ... | TRACE6
<Data>
Contains the data to transfer.
Analyzing Measurements (Basics)