General Instrument Setup
R&S
®
FSW
577
User Manual 1173.9411.02 ─ 19
A running Sequencer operation is aborted when you start a self-alignment.
Remote command:
Aborting the Self-alignment
As long as the self-alignment data is being collected the procedure can be cancelled
using the "Abort Self-alignment" button.
Starting the Touchscreen Alignment
Starts the touchscreen alignment.
Tap the 4 markers on the screen as you are asked to do. The touchscreen is aligned
according to the executed pointing operations.
Alignment Results
Information on whether the alignment was performed succesfully and on the applied
correction data is displayed. The results are available until the next self-alignment
process is started or the instrument is switched off.
Remote command:
9.3.4 System Configuration Settings
The system configuration information and settings are provided in the "System Config-
uration" dialog box which is displayed when you press the SETUP key and then select
"System Configuration".
............................................................................................577
Information on Versions and Options
....................................................................578
.................................................................................................579
.................................................................................................580
General Configuration Settings
.............................................................................581
9.3.4.1
Hardware Information
An overview of the installed hardware in your R&S
FSW is provided in the "Hardware
Info" tab of the "System Configuration" dialog box.
Every listed component is described by its serial number, order number, model infor-
mation, hardware code, and hardware revision.
This information can be useful when problems occur with the instrument and you
require support from Rohde & Schwarz.
General Instrument Settings