Dwell time
Consider the following when defining the dwell time:
●
Unknown signals
: select a dwell time of at least 1
second to ensure that pulses
down to a frequency of 5
Hz are weighted correctly
●
Pulsed signals or signals that fluctuate slowly
: the dwell time must cover at
least the time until the first signal peak is measured; can require long dwell time
●
unmodulated signals or signals with a high modulation frequency
: the dwell
time must cover at least the time until the first signal peak is measured; usually
shorter than for pulsed signals