Remote Commands
R&S
®
FSW
912
User Manual 1173.9411.02 ─ 19
[SENSe:]AVERage<n>[:STATe<t>]
<State>
This command turns averaging for a particular trace in a particular window on and off.
Parameters:
<State>
ON | OFF
Usage:
SCPI confirmed
[SENSe:]AVERage<n>:TYPE
<Mode>
This command selects the trace averaging mode.
Parameters:
<Mode>
VIDeo
The logarithmic power values are averaged.
LINear
The power values are averaged before they are converted to
logarithmic values.
POWer
The power level values are converted into unit Watt prior to
averaging. After the averaging, the data is converted back into
its original unit.
*RST:
VIDeo
Example:
AVER:TYPE LIN
Switches to linear average calculation.
Usage:
SCPI confirmed
Manual operation:
See
[SENSe:][WINDow<n>:]DETector<trace>[:FUNCtion]
<Detector>
Defines the trace detector to be used for trace analysis.
For details see
chapter 7.3.1.1, "Mapping Samples to Sweep Points with the Trace
For EMI measurements, the trace detector is used for the initial peak search only, not
for the final test. The detector for the final test is configured using
If the EMI (R&S
FSW-K54) measurement option is installed and the filter type "CISPR"
is selected, additional detectors are available, even if EMI measurement is not active.
For details see
chapter 5.13.3.2, "Detectors and Dwell Time"
Analyzing Measurements (Basics)