Common Analysis and Display Functions
R&S
®
FSW
507
User Manual 1173.9411.02 ─ 19
For spectrogram displays, define which frame the next minimum is to be searched in.
Remote command:
Center Frequency = Marker Frequency
Sets the center frequency to the selected marker or delta marker frequency. A peak
can thus be set as center frequency, for example to analyze it in detail with a smaller
span.
This function is not available for zero span measurements.
Remote command:
Reference Level = Marker Level
Sets the reference level to the selected marker level.
Remote command:
7.4.3 Marker Function Configuration
Special marker functions can be selected via the "Marker Function" dialog box.
To display this dialog box, do one of the following:
●
Press the MKR FUNC key, then select the "Select Marker Function" softkey.
●
In the "Overview", select "Analysis", and switch to the vertical "Marker Function
Config" tab.
The remote commands required to define these settings are described in
ter 11.8.3, "Working with Markers"
Precise Frequency (Signal Count) Marker
............................................................508
..................................................................................509
Phase Noise Measurement Marker
.......................................................................510
................................................................................................ 512
Marker Usage