Measurements
R&S
®
FSW
198
User Manual 1173.9411.02 ─ 19
Fig. 5-32: SEM results for multiple sub blocks
Retrieving Results via Remote Control
The measurement results of the spectrum emission mask
test
can be retrieved using
the
CALC:LIM:FAIL?
command from a remote computer; see
on page 987 for a detailed description.
The
power
result for the reference range can be queried using
CALC:MARK:FUNC:POW:RES? CPOW
;
The
peak
power for the reference range can be queried using
CALC:MARK:FUNC:POW:RES? PPOW
, see
The measured power
trace
can be queried using
TRAC:DATA?
and
TRAC:DATA:X?
,
The measured
peak power list
can be queried using
TRAC:DATA? LIST