Common Analysis and Display Functions
R&S
®
FSW
511
User Manual 1173.9411.02 ─ 19
To display the "Phase Noise Config" dialog box, do one of the following:
●
Press the MKR FUNC key, then select the "Select Marker Function" softkey. Then
select the "Phase Noise" button. Select the "Phase Noise Config" softkey.
●
In the "Overview", select "Analysis", and switch to the vertical "Marker Function
Config" tab. Then select the "Phase Noise" button. Select the "Phase Noise Con-
fig" softkey.
For more information see
chapter 7.4.1.8, "Measuring Phase Noise"
............................................................................... 511
......................................................................................... 511
Switching All Phase Noise Measurements Off
............................................................512
Phase Noise Measurement State
Activates or deactivates phase noise measurement for the reference point in the dia-
gram.
This function is only available for delta markers.
If activated, the delta markers display the phase noise measured at defined offsets
from the reference position.
Remote command:
Defining a Reference Point
Instead of using marker 1 as the reference marker, a fixed reference marker can be
defined for phase noise measurement.
Marker Usage