Remote Commands
R&S
®
FSW
1087
User Manual 1173.9411.02 ─ 19
Note that this command is maintained for compatibility reasons only. Use the
SYST:LANG
command for new remote control programs (see
Parameters:
<Mode>
DEFault | FSU | FSP | FSQ | FSV
Example:
SYST:COMP FSP
TRIGger[:SEQuence]:BBPower:HOLDoff
<Period>
This command defines the holding time before the baseband power trigger event.
The command requires the
Digital Baseband Interface (R&S
FSW-B17)
or the
Ana-
log Baseband Interface (R&S
FSW-B71)
.
Note that this command is maintained for compatibility reasons only. Use the
on page 851 command for new remote
control programs.
Parameters:
<Period>
Range:
150 ns to 1000 s
*RST:
150 ns
Example:
TRIG:SOUR BBP
Sets the baseband power trigger source.
TRIG:BBP:HOLD 200 ns
Sets the holding time to 200 ns.
TRIGger[:SEQuence]:RFPower:HOLDoff
<Time>
This command defines the holding time before the next trigger event. Note that this
command is available for any trigger source, not just RF Power.
Note that this command is maintained for compatibility reasons only. Use the
on page 851 command for new remote
control programs.
Parameters:
<Time>
Default unit: S
11.15 Programming Examples
Some advanced programming examples for complex measurement tasks are provided
here.
Programming Example: Performing a Basic Frequency Sweep